NIIN Output Data
NIIN:
NSN:
Item Name:
VISCOSIMETER SET
Reference/Part Number
Part Number | CAGE | Status | RNCC | RNVC | DAC | RNAAC | MSDS | SADC |
---|---|---|---|---|---|---|---|---|
PD-DRSAR-MAT-5-77 | 81996 | A | 3 | 2 | 5 | CT | ||
5911199 | 19204 | A | 3 | 2 | 1 | BF | ||
6630-0000-0000 | 99998 | A | 6 | 9 | 9 | PX | ||
PD-DRSAR-MAT-5-77 | 19204 | A | 3 | 2 | 3 | CT |
CAGE Information
Code | Company |
---|---|
19204 | ARMY UNITED STATES DEPARTMENT OF THE |
81996 | U S ARMY AVIATION AND MISSILE COMMAN |
99998 | DEFENSE SUPPLY CENTER RICHMOND INDUS |
Federal Supply Class
Title
CHEMICAL ANALYSIS INSTRUMENTS
Inclusions:
INCLUDES ACIDITY (PH) METERS; GAS ANALYZERS; ALAKALINITY MEASURING INSTRUMENTS; COLORIMETERS.
Exclusions:
EXCLUDES GAS DETECTING EQUIPMENT; MANUAL GAS ALARMS; CHEMICALS; IN VITRO DIAGNOSTIC SUBSTANCES AND REAGENTS.
Characteristics (Decoded)
MRC | Requirements Statement | Clear Text Reply |
---|---|---|
TEXT | GENERAL CHARACTERISTICS ITEM DESCRIPTION | THE VISCOSIMETER SET SHALL DETERMINE CHANGES IN THE KINEMATIC VISCOSITYWHICH AFFECTS OIL'S OVERALL CAPABILITYTO PROVIDE AN EFFECTIVE LUBRICATING MEDIUM 1 PER SYSTEM,THE VISCOMETER SHALL HAVE 6 RANGE'S 0-1 0-10 0-100 0-10000-10000 0-100000 CP SELECTABLE BY A SWITCH,DIGITAL READOUT METER,105-125 VOLTS 50-60 HZ,INSTANT READOUT WITH 1 PERCENT REPEATABILITY 2 PERCENT ACCURACY,1 1/4 INCH DIA STAINLESS STEEL HERMETICALLY SEALED TIP WITH -100 DEG TO 250 DEG F TEMPERATURE RANGE,CONTROLLED DRY BATH PLUS/MINUS 0.2 DEG C CONTROL TO 100 DEG C BY 10 POSITION SWITCH TO SELECT TEMPERATURE |