TUBE UNIT , MEDICAL X-RAY APPARATU

Definition:

CONSISTS OF A TUBE INSERT IN A SUITABLE HOUSING EQUIPPED WITH CABLE RECEPTACLES; WITHOUT CABLES.

CAGE Information

Code Company
81923 ARMY MEDICAL SERVICE C/O THE SURGEON

Federal Supply Class

Title

IMAGING EQUIPMENT AND SUPPLIES:

Inclusions:

INCLUDES MEDICAL X-RAY FILM; MEDICAL X-RAY FILM VIEWING EQUIPMENT; MEDICAL X-RAY FILM PROCESSING AND FINISHING EQUIPMENT AND SUPPLIES; X-RAY TUBES; COMPUTERIZED TOMOGRAPHY (CT) SCANNERS AND RELATED EQUIPMENT; MAGNETIC RESONANCE IMAGING MACHINE AND RELATED EQUIPMENT; POSITRON EMISSION TOMOGRAPHY (PET) SCANNERS AND RELATED EQUIPMENT; COMPUTED AND/OR DIRECT DIGITAL RADIOGRAPHY MACHINES AND RELATED EQUIPMENT; ULTRASOUND AND ECHOCARDIOGRAM SCANNERS AND RELATED EQUIPMENT; SCINTIGRAPHY EQUIPMENT; MEDICAL PICTURE ARCHIVING AND COMMUNICATION SYSTEMS (PACS) EQUIPMENT; MEDICAL THERMAL IMAGING EQUIPMENT; OPTICAL COHERENCE TOMOGRAPHY (OCT) SPECTROSCOPY MACHINES AND RELATED EQUIPMENT; AND IMAGING SUPPLIES AND CONTRAST DELIVERY SYSTEMS THAT ARE NOT USED BY OTHER MEDICAL DISCIPLINES.

Exclusions:

EXCLUDES INDUSTRIAL IMAGING EQUIPMENT (X-RAY, THERMAL AND THE REST;) SUPPLIES AND EQUIPMENT THAT CAN BE USED BY OTHER MEDICAL DISCIPLINES (SEE MEDICAL AND SURGICAL INSTRUMENTS, EQUIPMENT AND SUPPLIES AND HOSPITAL AND SURGICAL CLOTHING AND RELATED SPECIAL PURPOSE ITEMS.)

Characteristics (Decoded)

MRC Requirements Statement Clear Text Reply
ALQG CONTAINER TYPE ANY ACCEPTABLE
CXCY PART NAME ASSIGNED BY CONTROLLING AGENCY TUBE UNIT MED X-RAY APPAR TUNGSTEN-RHENIUM ALLOY 150 KILOVOLTS
CTGC ANODE HEAT CAPACITY RATING IN HEAT UNITS 300000
CLPP FOCAL SPOT QUANTITY 2
CLPJ TUBE INSERT ANODE TYPE STANDARD SPEED ROTATING
CMGS HOUSING X-RAY PROTECTIVE VALUE X-RAY PROTECTIVE VALUE OF HOUSING SHALL BE IN ACCORDANCE WITH DHEW PUBLICATION NO. FDA 75-8003 JUL 1974
CTBN PORT TO CABLE RECEPTACLE ANGLE IN DEG 270.0
CQWH TARGET ANGLE IN DEG 15.0
CLPQ FOCAL SPOT SIZE 0.6 MM SQ; 2.0 MM SQ
CQTT TARGET BASE MATERIAL MOLYBDENUM
CLPK TUBE INSERT FLUOROSCOPIC PEAK VOLTAGE RATING NOT RATED
CSJQ TARGET FACE MATERIAL TUNGSTEN-RHENIUM ALLOY
CLPL TUBE INSERT RADIOGRAPHIC PEAK VOLTAGE RATING 150.0 KILOVOLTS