National Stock Number (NSN) Output Data
NSN:
NIIN:
Item Name:
THERAPY UNIT , PULSED , ELECTROMAGNE
Reference/Part Number
Part Number | CAGE | Status | RNCC | RNVC | DAC | RNAAC | MSDS | SADC |
---|---|---|---|---|---|---|---|---|
D-103 | 042J2 | A | 3 | 2 | 4 | ZZ | ||
95-0802 | 89875 | A | 5 | 1 | 4 | ZZ |
CAGE Information
Code | Company |
---|---|
042J2 | DIAPULSE CORPORATION OF AMERICA |
89875 | DEFENSE LOGISTICS AGENCY DIRECTORATE |
Federal Supply Class
Title
HOSPITAL FURNITURE, EQUIPMENT, U
Inclusions:
INCLUDES ORTHOPEDIC EQUIPMENT; OPERATING LIGHTS; PHYSIOTHERAPY EQUIPMENT; STERILIZERS; WHEELCHAIRS; LITTERS; HOSPITAL BEDS; RESTRAINT EQUIPMENT.
Exclusions:
Characteristics (Decoded)
MRC | Requirements Statement | Clear Text Reply |
---|---|---|
SUCB | UNIT OF ISSUE CUBE | 10.0000 CUBIC FEET |
CXCY | PART NAME ASSIGNED BY CONTROLLING AGENCY | THERAPY UNIT PULSED ELECTROMAGNETIC 117V 60HZ AC 5 AMPS |
TEXT | GENERAL CHARACTERISTICS ITEM DESCRIPTION | DIAPULSE UNIT IS A NON-THERMAL, HIGH PEAK POWER, HIGH FREQUENCY SHORT WAVE, ELECTROMAGNETIC ENERGY DEVICE; INPUT POWER 117 VAC, 60 HZ, 5 AMPS; FREQUENCY 27.12 MHZ; PULSED POWER OUTPUT 293 TO 975 WATTS (MAX) IN SIX STEPS; PULSE DURATION 65 MICROSECONDS; 80 TO 600 PULSES PER SECOND IN SIX STEPS; FAIL SAFE CIRCUIT PROTECTION; HYBRID DESIGN; 21.5 IN. WIDE BY 19 IN. DEEP BY 44 IN. HIGH; BUILT-IN DIGITAL COMPLIANCE METER; BUILT-IN PULSE FREQUENCY METER; PUSH-BUTTON CONTROLLED; VARIABLE TUNING TREATMENT HEAD; FLEXIBLE ARM EXTENSION HEIGHT 6'6"/EXTENSION LENGTH 3'7"; SINGLE CONSOLE HEAVY GAUGE ALUMINUM ON CASTERS; DIAPULSE CALIBRATION METER INCLUDED; ACCELERATES TISSUE HEALING |
ECWT | EXTERIOR CONTAINER WEIGHT | 150.0000 POUNDS |