Part Number
Definition:
CAGE Information
Code | Company |
---|---|
81348 | FEDERAL SPECIFICATIONS PROMULGATED B |
Federal Supply Class
Title
IMAGING EQUIPMENT AND SUPPLIES:
Inclusions:
INCLUDES MEDICAL X-RAY FILM; MEDICAL X-RAY FILM VIEWING EQUIPMENT; MEDICAL X-RAY FILM PROCESSING AND FINISHING EQUIPMENT AND SUPPLIES; X-RAY TUBES; COMPUTERIZED TOMOGRAPHY (CT) SCANNERS AND RELATED EQUIPMENT; MAGNETIC RESONANCE IMAGING MACHINE AND RELATED EQUIPMENT; POSITRON EMISSION TOMOGRAPHY (PET) SCANNERS AND RELATED EQUIPMENT; COMPUTED AND/OR DIRECT DIGITAL RADIOGRAPHY MACHINES AND RELATED EQUIPMENT; ULTRASOUND AND ECHOCARDIOGRAM SCANNERS AND RELATED EQUIPMENT; SCINTIGRAPHY EQUIPMENT; MEDICAL PICTURE ARCHIVING AND COMMUNICATION SYSTEMS (PACS) EQUIPMENT; MEDICAL THERMAL IMAGING EQUIPMENT; OPTICAL COHERENCE TOMOGRAPHY (OCT) SPECTROSCOPY MACHINES AND RELATED EQUIPMENT; AND IMAGING SUPPLIES AND CONTRAST DELIVERY SYSTEMS THAT ARE NOT USED BY OTHER MEDICAL DISCIPLINES.
Exclusions:
EXCLUDES INDUSTRIAL IMAGING EQUIPMENT (X-RAY, THERMAL AND THE REST;) SUPPLIES AND EQUIPMENT THAT CAN BE USED BY OTHER MEDICAL DISCIPLINES (SEE MEDICAL AND SURGICAL INSTRUMENTS, EQUIPMENT AND SUPPLIES AND HOSPITAL AND SURGICAL CLOTHING AND RELATED SPECIAL PURPOSE ITEMS.)
Characteristics (Decoded)
MRC | Requirements Statement | Clear Text Reply |
---|---|---|
CMGQ | CASSETTE SIZE | 8 IN. BY 10 IN. |
FREQ | FREQUENCY IN HERTZ | 60.0 |
AKCV | DRIVE TYPE | MOTOR |
CLPX | RADIOGRAPHIC TIMER TYPE | EXPOSURE |
FEAT | SPECIAL FEATURES | REQUISITION SCHOULD STATE WHETHER HIGH-SPEED INCERTS ARE REQUIRED |
CLQF | TRANSFORMER OUTPUT CAPACITY | 500 MA AT 125 KV PEAK,300 MA AT 150 KV PEAK |
CCFN | SCREEN TYPE | W/SPOT-FILM |
BPKZ | RECTIFICATION TYPE | FULL-WAVE |
AAXX | MOUNTING TYPE | STATIONARY |
SUPP | SUPPLEMENTARY FEATURES | REQUISITION SHOULD STATE FLOOR-LOADING CAPACITY,CEILING STRENGTH AND HEIGHT,WALL AND FLOOR CONSTRUCTION (CONCRETE,WOOD,ETC),ELECTRICAL SUPPLY AND CAPACITY,DIAGRAM OF ROOM INDICATING DIMENSIONS,COMPATABILITY W/EXISTING FACILITIES,WHETHER OR NOT CONTRACTOR INSTALLATION SERVICE IS REQUIRED,AND WHETHER HIGH-SPEED INCERTS ARE REQUIRED |
CLPG | TUBE COOLING METHOD | AIR AND AIR |
CLQD | TRANSFORMER RECTIFICATION METHOD | ELECTRON TUBE |
CMGN | SCREEN LENGTH | 12.000 INCHES NOMINAL |
AKYD | ACCESSORY COMPONENTS AND QUANTITY | PHOTO-TIMING CIRCUIT 1,COMPRESSION DEVICE 1,MULTIDIAPHRAGM ADJUSTABLE RADIOGRAPHIC CONE 1,CONTROL /SEPARATE KILOVOLTAGE/FOR FLUOROSCOPIC KILOVO LTAGES 1,FOOTREST 1,MILLIAMPERE-SECOND METER 1,SPOT-FILM CHANGE-OVER DEVICE 1,MYLOGRAPHIC SUPPORT PATIENT 1,SHOULDER SUPPORT 1 |
CLPT | TILT RANGE | 90 DEG TRENDELENBURG TO VERTICAL |
CMGP | SCREEN WIDTH | 12.000 INCHES NOMINAL |
BNBB | TABLE TYPE | TILTING |
CLPW | BUCKY DIAPHRAGM | INCLUDED |
CTJL | FLUOROSCOPIC SCREEN TYPE | ANY ACCEPTABLE |
APGF | DESIGN TYPE | STATIONARY |
AQZF | CONTROL TYPE | VERTICAL |
AWGH | TUBE TYPE | DOUBLE FOCUS, ROTATING ANODE RADIOGRAPHIC AND DOUBLE FOCUS, ROTATING ANODE FLUOROSCOPIC |
APHE | OPERATION METHOD | AUTOMATIC |
CXCY | PART NAME ASSIGNED BY CONTROLLING AGENCY | X-RAY APPARATUS RADIOGR-FLUOROSCOPIC 500MA 150PKV 230V 60HZ AC |
ARQE | ELECTRIC POWER RATING | 65000.0 WATTS NOMINAL |
ELEC | VOLTAGE IN VOLTS | 190.0 AND 250.0 |
CLPY | FLUOROSCOPIC TIMER TYPE | SYNCHRONOUS MOTOR |
ARNF | TUBE QUANTITY | 1 AND 1 |
CLPH | TUBE MOUNTING TYPE | TUBE STAND AND TABLE |
FAAZ | PHASE | SINGLE |
EXQT | EXTERIOR CONTAINER QUANTITY (NON-CORE) | 1 |
CLPR | TUBE STAND MOUNTING TYPE | OVERHEAD |