Part Number
Part Number:
NSN:
NIIN:
Item Name:
CALIBRATOR , MULTIMETER
CAGE Information
Code | Company |
---|---|
98752 | WARNER ROBINS AIR LOGISTICS CENTER |
Federal Supply Class
Title
ELECTRICAL AND ELECTRONIC PROPER
Inclusions:
INCLUDES TEST LEADS AND TEST LEAD ATTACHMENTS; TEST INSTRUMENTS DESIGNED FOR COMMUNICATION EQUIPMENT; TEST INSTRUMENTS DESIGNED FOR USE WITH ELECTRONIC EQUIPMENT CLASSIFIED IN TWO OR MORE FSC GROUPS.
Exclusions:
EXCLUDES RADIAC EQUIPMENT
Characteristics (Decoded)
MRC | Requirements Statement | Clear Text Reply |
---|---|---|
FTLD | FUNCTIONAL DESCRIPTION | HIGH ACCURACY DC,AC AND RESISTANCE CALIBRATION SYSTEM, THAT MAY INCLUDE A MONOLITHIC INTEGRATED UNTI OR SEPERATE AMPLIFIER AS NECESSARY |
FEAT | SPECIAL FEATURES | UNIT SHALL MEET ALL ACCURACY REQUIREMENTS WHEN OPERATING TEMPERATURE ISBETWEEN 0 DEGREES CELSIUS AND 50 DEGREES CELSIUS. WARM UP TIME SHALL NOT EXCEED 30 MINUTES. DC VOLTAGE BETTER THAN0 TO 1100 VOLTS, OUTPUT VOLTAGE WITHIN+ OR - 5 DEGREES CELSIUS OF THE CALIBRATION TEMPERATURE. SHALL BE CAPABLE OFACCURACY SPECIFICATIONS AT ANY CALIBRATION TEMPERATURE BETWEEN 15 AND 35 DEGREES CELSIUS. AC VOLTAGE, .22MV TO 1100V, ACCURACY .0022 TO 1100 VOLTS, FREQUENCY RANGE 40 - 300 KHZ. WEIGHT SHALLNOT EXCEED 75 POUNDS,7"H X 22.75"D(SHALL NOT EXCEED) SHALL FIT IN A STANDARD 19" RACK. POWER CONSUMPTION, LESS THAN 300 WATTS |