Part Number
Definition:
CONSISTS OF A TUBE INSERT IN A SUITABLE HOUSING EQUIPPED WITH CABLE RECEPTACLES; WITHOUT CABLES.
CAGE Information
Code | Company |
---|---|
37480 | RAYTHEON CO MACHLETT LABORATORIES IN |
Federal Supply Class
Title
IMAGING EQUIPMENT AND SUPPLIES:
Inclusions:
INCLUDES MEDICAL X-RAY FILM; MEDICAL X-RAY FILM VIEWING EQUIPMENT; MEDICAL X-RAY FILM PROCESSING AND FINISHING EQUIPMENT AND SUPPLIES; X-RAY TUBES; COMPUTERIZED TOMOGRAPHY (CT) SCANNERS AND RELATED EQUIPMENT; MAGNETIC RESONANCE IMAGING MACHINE AND RELATED EQUIPMENT; POSITRON EMISSION TOMOGRAPHY (PET) SCANNERS AND RELATED EQUIPMENT; COMPUTED AND/OR DIRECT DIGITAL RADIOGRAPHY MACHINES AND RELATED EQUIPMENT; ULTRASOUND AND ECHOCARDIOGRAM SCANNERS AND RELATED EQUIPMENT; SCINTIGRAPHY EQUIPMENT; MEDICAL PICTURE ARCHIVING AND COMMUNICATION SYSTEMS (PACS) EQUIPMENT; MEDICAL THERMAL IMAGING EQUIPMENT; OPTICAL COHERENCE TOMOGRAPHY (OCT) SPECTROSCOPY MACHINES AND RELATED EQUIPMENT; AND IMAGING SUPPLIES AND CONTRAST DELIVERY SYSTEMS THAT ARE NOT USED BY OTHER MEDICAL DISCIPLINES.
Exclusions:
EXCLUDES INDUSTRIAL IMAGING EQUIPMENT (X-RAY, THERMAL AND THE REST;) SUPPLIES AND EQUIPMENT THAT CAN BE USED BY OTHER MEDICAL DISCIPLINES (SEE MEDICAL AND SURGICAL INSTRUMENTS, EQUIPMENT AND SUPPLIES AND HOSPITAL AND SURGICAL CLOTHING AND RELATED SPECIAL PURPOSE ITEMS.)
Characteristics (Decoded)
MRC | Requirements Statement | Clear Text Reply |
---|---|---|
ALQG | CONTAINER TYPE | ANY ACCEPTABLE |
CTGC | ANODE HEAT CAPACITY RATING IN HEAT UNITS | 300000 |
CLPJ | TUBE INSERT ANODE TYPE | HIGH SPEED ROTATING |
CLPP | FOCAL SPOT QUANTITY | 2 |
CQWH | TARGET ANGLE IN DEG | 12.0 |
CMGS | HOUSING X-RAY PROTECTIVE VALUE | X-RAY PROTECTIVE VALUE OF HOUSING SHALL BE IN ACCORDANCE WITH DHEW PUBLICATION NO. FDA 75-8003 JUL 1974 |
CLPQ | FOCAL SPOT SIZE | 0.6 MM SQ; 1.2 MM SQ |
CTBN | PORT TO CABLE RECEPTACLE ANGLE IN DEG | 90.0 |
CBBL | FEATURES PROVIDED | GRID CONTROL AND HIGH EMISSION |
CQTT | TARGET BASE MATERIAL | MOLYBDENUM |
CLPK | TUBE INSERT FLUOROSCOPIC PEAK VOLTAGE RATING | NOT RATED |
CSJQ | TARGET FACE MATERIAL | TUNGSTEN-RHENIUM ALLOY |
CXCY | PART NAME ASSIGNED BY CONTROLLING AGENCY | TUBE UNIT MED X-RAY APPAR HI-SPEED TUNGSTEN-RHENIUM 150 KILOVOLT |
CLPL | TUBE INSERT RADIOGRAPHIC PEAK VOLTAGE RATING | 150.0 KILOVOLTS |